N2 - Objective This study aimed to ascertain the accuracy of measure arterial stiffness using the HUAWEI GT 3 Pro smartwatch and pOpmètre device against the SphygmoCor (algorithms: intersect tangent and maximum of the second derivate). Methods Twenty-three physically active postmenopausal women (age: 58.9 ± 3.2 years; body mass index: 26.3 ± 4.8 kg/m2) were recruited. Carotid-femoral pulse wave velocity, finger-toe pulse wave velocity, and wrist-finger pulse wave velocity were obtained using SphygmoCor, pOpmètre and HUAWEI GT 3 Pro devices in a randomized order. Additionally, the pulse mean carotid-femoral and finger-toe pulse transit time was registered for SphygmoCor and pOpmètre, respectively. Results Lower values of pulse wave velocity were recorded by HUAWEI in comparison with SphygmoCor with both algorithms, whereas no significant differences were detected between SphygmoCor and pOpmètre results. Pulse wave velocity values from SphygmoCor were positively correlated with pOpmètre results (r = 0.464 and r = 0.451 using intersect tangent and second derivative algorithms), whereas this was not the case with those obtained from HUAWEI. Coefficients of bias of Lin's concordance coefficients close to 1 (0.832 and 0.831 for intersect tangent and second derivative algorithm, respectively) and mean bias close to 0 from Bland-Altman analysis suggested an acceptable agreement between pulse wave velocity obtained from SphygmoCor and pOpmètre. Conclusions Our results suggest an acceptable concordance of pulse wave velocity values recoded by SphygmoCor and pOpmètre, whereas this was not the case for data obtained from HUAWEI GT 3 Pro smartwatch. Therefore, the pOpmètre may be a viable alternative for assessing arterial stiffness, but measurement via the smartwatch device cannot be recommended.
AB - Objective This study aimed to ascertain the accuracy of measure arterial stiffness using the HUAWEI GT 3 Pro smartwatch and pOpmètre device against the SphygmoCor (algorithms: intersect tangent and maximum of the second derivate). Methods Twenty-three physically active postmenopausal women (age: 58.9 ± 3.2 years; body mass index: 26.3 ± 4.8 kg/m2) were recruited. Carotid-femoral pulse wave velocity, finger-toe pulse wave velocity, and wrist-finger pulse wave velocity were obtained using SphygmoCor, pOpmètre and HUAWEI GT 3 Pro devices in a randomized order. Additionally, the pulse mean carotid-femoral and finger-toe pulse transit time was registered for SphygmoCor and pOpmètre, respectively. Results Lower values of pulse wave velocity were recorded by HUAWEI in comparison with SphygmoCor with both algorithms, whereas no significant differences were detected between SphygmoCor and pOpmètre results. Pulse wave velocity values from SphygmoCor were positively correlated with pOpmètre results (r = 0.464 and r = 0.451 using intersect tangent and second derivative algorithms), whereas this was not the case with those obtained from HUAWEI. Coefficients of bias of Lin's concordance coefficients close to 1 (0.832 and 0.831 for intersect tangent and second derivative algorithm, respectively) and mean bias close to 0 from Bland-Altman analysis suggested an acceptable agreement between pulse wave velocity obtained from SphygmoCor and pOpmètre. Conclusions Our results suggest an acceptable concordance of pulse wave velocity values recoded by SphygmoCor and pOpmètre, whereas this was not the case for data obtained from HUAWEI GT 3 Pro smartwatch. Therefore, the pOpmètre may be a viable alternative for assessing arterial stiffness, but measurement via the smartwatch device cannot be recommended.
KW - Algorithms
KW - Female
KW - Humans
KW - Middle Aged
KW - Postmenopause/physiology
KW - Pulse Wave Analysis/instrumentation
KW - Vascular Stiffness/physiology
UR - http://www.scopus.com/inward/record.url?scp=85199814853&partnerID=8YFLogxK
U2 - 10.1097/gme.0000000000002383
DO - 10.1097/gme.0000000000002383
M3 - Article
C2 - 38916283
AN - SCOPUS:85199814853
SN - 1072-3714
VL - 31
SP - 709
EP - 715
JO - Menopause
JF - Menopause
IS - 8
ER -